Art No. |
JSCO-150T |
JSCO-300T |
JSCO-1200T |
|
Chamber Volume (Capacity) |
168 L |
336 L |
1200 L |
|
Dimensions |
Inner |
600 x 560 x 500 |
600 x 560 x 1000 |
1000 x 1000 x 1200 |
W x D x H mm |
Outer |
845 x 1100 x 1240 |
845 x 1100 x 1740 |
1245 x 1540 x 1940 |
|
Clearance |
845 x 1110 x 1250 |
845 x 1110 x 1750 |
1245 x 1550 x 1950 |
Temperature Range |
ambient +10 oC to 250 oC |
|||
Accuracy at 120 oC |
±0.3 oC |
±0.3 oC |
±0.5 oC |
|
Uniformity at 120 oC |
±5.0 oC |
±5.0 oC |
±8.0 oC |
|
HEPA Filter |
H14 (ISO Class 5) HI-TEMP HEPA Filter Efficiency 99.995%>0.3um particles |
|||
No. of Shelves / Max. No. |
2 / 14 |
3 / 36 |
4 / 36 |
|
No. of Ventilation damper |
1 |
|||
Heater Capacity |
2 x 1.0 kW |
2 x 2.0 kW |
3 x 2.5 kW |
|
Electrical Requirement |
220VAC 50/60Hz 1P |
220VAC 50/60Hz 1P |
220VAC 1P or 380VAC 3P 4W |
|
Power Rating |
11.0 Amp |
20.2 Amp |
- |
|
Power Plug |
CEE 7/7 Schuko |
30 AMP |